2-HA bioceramic coating"/> 2 and TiO2-HA bioceramic coatings, were prepared by micro-arc oxidation and micro-arc oxidation-hydro-thermal methods. Determination of the hemocompatibility of the coatings through a hemolysis experiment, dynamic clotting time test, and platelet adhesion experiment was also performed. The results demonstrated that: Hemolysis rates of all samples were less than 5%, no hemolysis was observed. Compared with titanium and TiO2 bioceramic coating of titanium, TiO2-HA bioceramic coating of titanium exhibited lower hemolysis rate, slower change in dynamic clotting time and less platelet adherence. The proposed TiO2-HA bioceramic coating further showed a lower degree of deformation, better inhibition of platelet aggregation and deformation, and excellent hemocompatibility."/>
School of Mechanical Engineering,Jiangsu University,,School of Mechanical Engineering,Jiangsu University,Laboratory Animal Research CenterS,Jiangsu University,Laboratory Animal Research CenterS,Jiangsu University
R138
[Xu Lin, Ding Jianning, Xu Xiaojing, HeYuanqing, Lei Xiaochun. Blood Compatibility of TiO2-HA Bioceramic Coating on Titanium[J]. Rare Metal Materials and Engineering,2017,46(5):1299~1304.]